Introduction
Phosphatidylethanols (PEth) serve as highly specific biomarkers for alcohol intake, as these abnormal phospholipids are synthesized in the body only in the presence of ethanol. In contrast to Ethyl Glucuronide (EtG), with its short urinary detection window of a few days, and Carbohydrate-Deficient Transferrin (CDT), which mainly reflects chronic alcohol use, PEth provides as a marker over an extended period.
While liquid chromatography tandem mass spectrometry (LC-MS/MS) can identify various PEth homologs, clinical testing typically targets the most prevalent forms, specifically PEth 16:0/18:1 and 16:0/18:2. Because PEth accumulates in erythrocytes and has a slow elimination rate, it can provide a detection window of several weeks. Furthermore, the faster synthesis rate and high abundance of PEth 16:0/18:2 enhance sensitivity for detecting recent alcohol consumption. Here, we describe a fast and robust method for the quantification of PEth 16:0/18:1 and 16:0/18:2 in whole blood.
Sample preparation is achieved through extraction of analytes and the use of dedicated internal standards for all PEth homologs. The 6PLUS1® whole blood calibrator set covers the relevant calibration ranges for PEth and meets both forensic and clinical calibration requirements. Quality control is supported by the use of our two MassCheck® controls, which are designed to reflect the relevant decision limits.
Material & Methods
Sample Preparation
For sample preparation, 20 µL of research sample/calibrator/MassCheck® control is pipetted into a 1.5 mL reaction vial. Subsequently, 150 μL of Internal Standard Mix (63004) is added and vortexed for at least 30 seconds, followed by centrifugation for 5 minutes at 15,000 g. 80 µL of the resulting supernatant is transferred into an autosampler vial.
LC-MS/MS
Injection volume of 1–20 μL is used. Chromatographic separation is carried out with two Mobile Phases at a flow rate of 0.4 mL/min using Mobile Phases A and B (order nos. 63001, 63002) and an Analytical Column (order no. 63100).
Electrospray ionization in negative ion mode is used. The multiple reaction monitoring (MRM) transitions used for quantification are listed in Table 1.
Quantification
A full calibration of the analysis system for each series of measurements is performed with 6PLUS1® multi level calibrator (order no. 63039). Calibration curves are constructed by calculating the analyte to internal standard (ISTD) peak area ratio on the y axis against calibrator concentrations on the x axis. Then a calibration curve is plotted for both PEth homolouges using linear regression and 1/x²︎ weighting.


Table 1: Multiple reaction monitoring transitions of Phosphatidylethanol 16:0/18:1 and 16:0/18:2, together with their respective internal standards


Table 2: Interpretation of Phosphatidylethanol 16:0/18:1 values
Results
The run time is 5 min (see chromatogram in fig. 1). The calibration range of 10-1000 ng/mL covers the forensic and clinical relevant concentrations (table 2).
Conclusion
This page describes a protocol for the determination of PEth 16:0/18:1 and PEth 16:0/18:2 in whole blood using Chromsystems components.
Outlook
Further research will focus on developing a protocol for automated sample preparation using 96-well plates and liquid handling systems such as the Chromsystems MassPrep®.


Fig. 1: Chromatogram of phosphatidylethanol 16:0/18:1 and 16:0/18:2 at a concentration of 300 ng/mL each, with their respective internal standards


Fig. 2: Chemical structure of Phosphatidylethanol 16:0/18:1 and 16:0/18:2
The following products have been released and can be ordered (More are to come)


Regulatory status: For research use only (RUO).
Further products are currently being developed. Please contact your local reprensentative for further information
| Product | Order no. |
| Mobile Phase | 63001 |
| Rinsing Solution | 63009 |
| Internal Standard Mix | 63004 |
| Extraction Reagent | 63005 |
| Tuning Mix, Analytes and Internal Standards | 63015 |
| Analytical Column | 63100 |
| Product | Order no. |
| 6PLUS1® Multilevel Whole Blood Calibrator Set MassChrom® Phosphatidylethanol (PEth) |
63039 |
| MassCheck® Phosphatidylethanol (PEth) Whole Blood Control Set Level I |
6310 |
| MassCheck® Phosphatidylethanol (PEth) Whole Blood Control Set Level II |
6320 |