Sample Prep Set VMA, HVA, 5-HIAA with Collection Plates

Order No.: 80800/96, for 96 tests
Parameters:
5-Hydroxyindoleacetic Acid (5-HIAA), Homovanillic Acid (HVA), Vanillylmandelic Acid (VMA)

Simple and fast sample prep
Isotopically labelled internal standard for each analyte
6PLUS1® Multilevel Calibrator Set and MassCheck® controls
Fully automated workflow solution available

In addition to a Sample Prep Set the products of the Chromatographic Platform are required.

5-Hydroxyindoleacetic Acid (5-HIAA)

Homovanillic Acid (HVA)

Vanillylmandelic Acid (VMA)

MassChrom® Biogenic Amines/Metabolites in Urine - LC-MS/MS
With collection plates for 96 tests (80800/96) or 480 tests (80800/480), suitable for manual and automated sample preparation

 

Clinical Relevance

This Sample Prep Set allows the quantitative determination of vanillylmandelic acid (VMA), homovanillic acid (HVA), 5-hydroxyindoleacetic acid (5-HIAA) in patient urine samples by LC-MS/MS. It is used for the diagnostics of neuroblastoma and carcinoid tumours.

 

Product advantage

  • Simple and fast sample prep consisting of 2 pipetting steps and subsequent injection
  • Isotopically labelled internal standard for each analyte
  • 6PLUS1® Multilevel Calibrator Set and MassCheck® controls ensure high precision of results
  • Fully automated workflow solution available

This Sample Prep Set encompasses all required components for an efficient and fast sample preparation.

A fully automated workflow solution is also available.

Alternatively, there is also a method available using autosampler vials (80800).

 

More Information
Method of Analysis LC-MS/MS
Number of Tests 96
Limit of quantification 0.2 - 2.6 mg/l
Linearity

VMA, 5-HIAA up to 150 mg/l
HVA up to 250 mg/l

Recovery 95 - 113 %
Intraassay CV 2.5 - 6.7 %
Interassay CV 3.7 - 7.7 %
Analysis Time 3 min
Specimen 24h urine or spontaneous urine
Pre-analytic Treatment

24 h urine is collected in a suitable container with 10 ml of 10–25 % HCl (VMA, HVA) or 10 ml glacial acetic acid (5-HIAA).
Spontaneous urine can be also analysed (see instruction manual).

Patient specimens are stable for up to 14 days at +2 to +8 °C in amber coloured vials.
For longer storage periods keep samples frozen below -18 °C.

Sample Preparation
  • Pipette 50 μl sample/calibrator/MassCheck® control into a well of the collection plate.
  • Add 25 μl Internal Standard Mix.
  • Add 425 μl Stabilisation Buffer.
  • Seal collection plate with adhesive seal.
  • Agitate collection plate for 2 min at 1200 rpm (orbit 2 mm) and transfer to autosampler.
  • Inject eluates into the LC-MS/MS system.
Sample Stability The prepared samples are stable for up to 5 days (VMA and HVA) or up to 3 days ( 5-HIAA) at +2°C to +8°C in amber coloured vials. For longer storage periods keep samples frozen below -18°C.
Injection Volume ≤ 2 µl
Column temperature +25 °C
Ionisation ESI positive
MS/MS-Mode MRM