VMA, HVA, 5-HIAA Urine Calibration Standard
VMA, HVA, 5-HIAA Urine Calibration Standard, lyophilised
This lyophilised calibrator from Chromsystems is based on human urine and should be used according to the instruction manual. It is designed to calibrate the measurement method in clinical diagnostics for the quantitative determination of VMA, HVA, 5-HIAA in urine.
Content | 5 x 5 ml (lyoph.) | ||||||||
---|---|---|---|---|---|---|---|---|---|
Concentrations | Please note: All concentrations shown here are exemplatory. They do not reflect real data for the described products. Actual concentrations depend on the production lot of each product and can exclusively be found in the product information leaflet. Please ask for these product information leaflets or download them from the service site of the Chromsystems website.
|
||||||||
Reconstitution | Carefully remove the rubber stopper from the vial (Caution when removing the metal seal: Risk of injuries due to sharp cutting edges!). Add exactly 5.0 ml of distilled water into the vial. Replace the stopper and allow to sit for 10 to 15 minutes at room temperature. Swirl the vial to dissolve the contents until homogeneity. | ||||||||
Storage conditions | Stored at +2 to +8 °C unopened calibration standards are stable until the expiry date which is stated on the vial labels. The reconstituted calibration standard should be stored tightly capped at +2 to +8 °C. Generally the calibration standard can be stored for two days at +2 to +8 °C. If the calibration standard cannot be analysed within this period, it should be aliquoted and stored frozen below –18 °C (up to 3 months). |
||||||||
Method of Analysis | HPLC | ||||||||
Parameter | 5-Hydroxyindoleacetic Acid (5-HIAA), Homovanillic Acid (HVA), Vanillylmandelic Acid (VMA) |